Loop Test With User Defined Loop Current Value - Siemens SITRANS TS Series Operating Instructions Manual

Table of Contents

Advertisement

3. Use the
The loop test starts:
– The "EDIT" symbol flashes.
– The "Function check" symbol is displayed.
– The "Co" symbol (constant current mode) is displayed.
4. Change the preset value with the
5. Use the
The loop test starts.
6. End the loop test with the
8.3.12.2

Loop test with user defined loop current value

1. Navigate into the parameter view.
Navigating in the views (Page 60)
2. In the parameter view, select the "Loop test" parameter.
3. Use the
The loop test starts:
– The "EDIT" symbol flashes.
– The "Function check" symbol is displayed.
– The "Co" symbol (constant current mode) is displayed.
4. Change to "USER" with the
5. Use the
6. Set a value between 3.6 mA and 22.8 mA using the buttons
7. Use the
The loop test starts.
8. End the loop test with the
SITRANS TS100/TS200/TS300/TS500/TSinsert/TSthermowell
Operating Instructions, 08/2020, A5E47810090-AA
button to confirm.
button to confirm.
button.
button to confirm.
or
button to confirm.
button to confirm.
button.
8.3 Parameter assignment over device with display
or
button.
button.
Parameter assignment
or
.
87