Pump Pressure Adjust Regulator

The PUMP PRESSURE ADJUST regulator is used to control the maximum pressure of the
hydraulic pump. Turning the regulator knob in (clockwise) will increase the maximum pump
pressure. Turning the regulator knob out (counterclockwise) will decrease the maximum
pump pressure. When the PRESSURIZE CYLINDER switch is in the ON position, the
PUMP PRESSURIZE ADJUST regulator may be used to maintain cylinder pressure above a
certain level. The PUMP PRESSURE ADJUST may also be used to slowly and quietly
increase the cylinder pressure. This is done by turning the PUMP PRESSURE ADJUST
knob clockwise while the PRESSURIZE CYLINDER switch is in the ON position.

Temperature Limit Alarm

The TEMPERATURE LIMIT ALARM switch light will come on if the maximum allowable
cylinder temperature is exceeded. When a TEMPERATURE LIMIT ALARM occurs, the
heaters are automatically switched off. The TEMPERATURE LIMIT ALARM switch will
not go off and the heaters will not operate until the manual TEMPERATURE LIMIT
ALARM switch is pressed.

Temperature Limit Alarm Reset

The Temperature Limit Alarm Reset is used to reset the TEMPERATURE LIMIT ALARM.
When the alarm is tripped, the light will not go off and the heaters will not operate until the
TEMPERATURE LIMIT ALARM switch is pressed.