RF ATTENUATION CLAMP
KEMA 801A
USER MANUAL

Advertisement

Table of Contents
loading

Summary of Contents for Ametek TESEQ KEMA 801A

  • Page 1 RF ATTENUATION CLAMP KEMA 801A USER MANUAL...
  • Page 2 RF ATTENUATION CLAMP KEMA 801A USER MANUAL...
  • Page 3: Table Of Contents

    CONTENT Safety advice ........................5 1.1. General ..............................5 1.2. Installation ..............................5 1.3. Applicable safety standards ........................5 Unpacking, storage and transport ..................6 2.1. General ..............................6 2.2. Storage and transport ..........................6 2.3. Unpacking ..............................6 2.4. Scope of delivery and order information ....................
  • Page 4 KEMA 801A...
  • Page 5: Safety Advice

    1. SAFETY ADVICE The clamp can be part of a setup which works at dangerous voltages. WARNING: Improper or careless handling can be fatal! Use of the clamp is restricted to authorized and trained specialists These operating instructions form an integral part of the equipment and must be available to the operating personnel at all times. All the safety instructions and advice notes are to be observed. Neither Teseq GmbH nor any of the subsidiary sales organizations or the manufacturer can accept any responsibility for personal, material or consequential injury, loss or damage that results from improper use of the equipment and acces- sories.
  • Page 6: Unpacking, Storage And Transport

    2. UNPACKING, STORAGE AND TRANSPORT 2.1. General Save all packing materials! They will be needed in order to safely package the equipment for calibration service or repair Packaging materials Carton: Cardboard ■ Padding: CFC-free polystyrene foam ■ Plastic bags: Polyethylene ■...
  • Page 7: Application And Standard Requirements

    3. APPLICATION AND STANDARD REQUIREMENTS 3.1. General The KEMA 801A is recommended as an additional decoupling network (ferrite tube clamp) for immunity test according to IEC / EN 61000-4-6 when using the clamp injection method. It shall be inserted on all the cables between EUT and AE except the cable under test.
  • Page 8: Verification Of The Insertion Loss

    4. VERIFICATION OF THE INSERTION LOSS The insertion loss is measured in a 50 Ω system with CAL KEMZ, 10 dB attenuators, network analyzer and cables. In a first step the measuring set-up is normalized with two connected impedance measurement adapt- ers and 10 dB attenuators as shown in the figure below.
  • Page 9: Construction Of The Product

    5. CONSTRUCTION OF THE PRODUCT Mechanical interlock (adjustable) Ferrite cores, spring-loaded Figure 4: View to the opened KEMA 801A Figure 5: Side view 85-238526 E02...
  • Page 10: Technical Specification

    6. TECHNICAL SPECIFICATION Frequency range: 150 kHz to 1000 MHz Insertion loss: see graph Maximum cable diameter: 20 mm Dimension (LxWxH): 670 mm x 100 mm x 79 mm Weight: approx. 5.5 kg Environmental classification: Indoor use only Operating temperature: +5°C to +40 °C Relative humidity: up to 80%...
  • Page 11: Maintenance

    7. MAINTENANCE 7.1. General The KEMA 801A needs no special maintenance. The maintenance is limited to the cleaning of the ferrite contacts. No modifications are to be carried out on the KEMA 801A and accessories by the user. 7.2. Cleaning The cleaning shall be done with dry cloth.
  • Page 12 T +8610 8526 2111 F +8610 8526 2141 info.france@ametek.com chinasales@teseq.com Japan Germany AMETEK Co., Ltd. Nagoya Office Teseq GmbH T + 81 52 709 5501 T + 49 30 5659 8835 F + 81 52 709 5502 F + 49 30 5659 8834 cts-japan.sales@ametek.co.jp...