Section 2 - Curing Chamber Controls; Safety Procedures; Power Switch - Ametek CHANDLER ENGINEERING 7360V Instruction Manual

Vertical curing chamber
Table of Contents

Advertisement

Section 2 - Curing Chamber Controls

Safety Procedures

The following safety procedures are recommended to help ensure operator safety:
• Read and understand the instructions before operating the instrument.
• Never exceed the temperature or pressure ratings listed on the front panel.
• Disconnect main power before servicing the instrument.
• Turn off heater at the completion of the test.
• Do not attempt to open the CYLINDER DRAIN valve until the water temperature is
below 200°F (93°C).
• Use caution when opening the CYLINDER DRAIN valve at high temperatures and
pressures.
• Make certain that the side panels are in place when operating the instrument.
• Always install a new plug O-ring if the cylinder temperature has exceeded 350°F/177°C.
Control Panel Functions
Figure 2 shows how the front control panel is designed. You may find it convenient to refer
to this figure as you study the control panel functions.

Power Switch

The Power switch controls all electrical power to the instrument.
Temperature Controller
The Temperature Controller controls the 3000-Watt cylinder heater. The water temperature
is measured using a thermocouple inserted through the bottom of the cylinder. The controller
may be programmed in a variety of ways to provide different test conditions.
For the heater to function, the HEATER switch must be in the ON position and the
temperature controller must be operating. If a heat-up cycle must be stopped, the heater can
be turned OFF by switching the HEATER switch to the OFF position or by pressing the F1
(Auto / Manual) button on the controller.
SECTION 2 – CURING CHAMBER CONTROLS
2-1