9 Diagnosis, asset management and service
Failure
Code
Text message
F013
No valid measured value
available
F017
Adjustment span too
small
F025
Error in the lineariza-
tion table
F036
no operable sensor soft-
ware
F040
Error in the electronics
F041
Communication error
F042
Communication error
Secondary sensor
F080
General software error
F105
Measured value is de-
termined
F113
Communication error
F260
Error in the calibration
F261
Error in the instrument
settings
F264
Installation/Setup error
48
Cause
Gauge pressure or low pressure
Measuring cell defective
Adjustment not within specifi-
cation
Index markers are not continu-
ously rising, for example illogical
value pairs
Failed or interrupted software
update
Hardware defect
No connection to the sensor
electronics
No connection to the Second-
ary sensor
General software error
The instrument is still in the
switch-on phase, the measured
value could not yet be deter-
mined
Error in the internal instrument
communication
Error in the calibration carried out
in the factory
Error in the EEPROM
Error during setup
Error when carrying out a reset
Inconsistent settings (e.g.: dis-
tance, adjustment units with
application process pressure) for
selected application
Invalid sensor configuration (e.g.:
application electronic differential
pressure with connected differen-
tial pressure measuring cell)
Rectification
Exchange measuring cell
Send instrument for repair
Change the adjustment accord-
ing to the limit values
Check linearization table
Delete table/Create new
Repeat software update
Check electronics version
Exchanging the electronics
Send instrument for repair
Exchanging the electronics
Send instrument for repair
Check connection between sen-
sor and main electronics (with
separate version)
Check connection between Pri-
mary and Secondary sensor
Disconnect operating voltage
briefly
Wait for the end of the switch-
on phase
Disconnect operating voltage
briefly
Send instrument for repair
Exchanging the electronics
Send instrument for repair
> Repeat setup
Repeat reset
Modify settings
Modify connected sensor config-
uration or application
VEGABAR 86 • Foundation Fieldbus
DevSpec
Diagnosis Bits
Bit 0
Bit 1
Bit 2
Bit 3
Bit 4
Bit 13
Bit 28 of
Byte 0 ... 5
Bit 5
Bit 6
Bit 12
Bit 8
Bit 9
Bit 10