Appendix Aspecifications - Elitech ChloroChek Chloridometer 3400 User Manual

Table of Contents

Advertisement

APPENDIX A
SPECIFICATIONS
ChloroChek Chloridometer Model 3400
Sample Volume ......................................................................... 10 μL
Measurement Duration ............................................................. Approximately 20 Seconds
Reproducibility* ........................................................................ CV of 1.02% at 100 mmol/L
Measurement Display ............................................................... 0 - 999 mmol/L
Linearity Measuring Range ....................................................... 10 - 160 mmol/L
Resolution .................................................................................. 1 mmol/L over the entire measurement range
Integrated Stir Bar .................................................................... PTFE coated magnetic cylindrical stir bar
Data Output ............................................................................... Serial RS232/USB
Environment operating temperature ........................................ Indoor use; no direct sunlight
Operating temperature (ambient) ............................................. 50 ˚F to 95 ˚F / 10˚C to 35 ˚C
Storage temperature ................................................................ -40 ˚F to 158 ˚F / -40 ˚C to 70 ˚C
Room humidity ......................................................................... 5-80% (non-condensing)
Altitude ...................................................................................... up to 2000 m /6562 feet
Electrical Voltage ....................................................................... 100-120 VAC / 200-240 VAC
Frequency .................................................................................. 50-60 Hz
Power ......................................................................................... 20 VA
Electrical connection ................................................................. Detachable power supply cord
Fuses (2): HBC breaking capacity 1500 A ................................... 1.0 A for 100-120 VAC
0.5 A for 200-240 VAC
Memory backup ......................................................................... integral lithium cell: 10 years typical
Dimensions: (W x D x H) ........................................................... 20.5 x 22.0 x 36.0 cm (8.0 x 8.1 x 14.2 inches)
Weight ...................................................................................... Approximately 5.7 kg (12.54 lbs.)
*Generally limited by the sample dispensing system (pipette accuracy)