SECTION 3
OPERATING THE CHLOROCHEK

3.8 Ending a Measurement Series

1
Once all samples are measured, press STANDBY and then CANCEL. This will end the measurement
sequence.
NOTE:
Users can run up to 50 samples in each series before needing a fresh Working Solution. A freshly prepared
Working Solution should be a transparent red color. As a Working Solution is used it will appear cloudy. If a
freshly prepared Working Solution appears cloudy it should be discarded. The titration beaker should then be
carefully cleaned and a new Working Solution prepared.
2
Raise the handle to lift the electrode set out of the Working Solution.
3
Remove the stir bar from the Working Solution by using the Magnetic Stir Bar Retriever (RP-486).