Measurement
A measurement or mismeasurement is described in a ticket in an XML tag called MEASUREMENT, which contains
the following values:
BatchID
Batch ID appears if entered manually or set to automatic. A minus sign "-" is displayed if Batch
ID type is none, Single Sample is selected during measurement, or if it is a mismeasurement.
SampleID
Sample ID appears if entered manually or set to automatic. Void if Sample ID type is none and
will not be present if it is a mismeasurement.
DateTime
Date and time in combined ISO 8601 format (e.g. 2015-12-31T13:45).
Value
Measurement result with floating point number with a point as decimal separator (regardless
of language setting). Value will not be present if it is a mismeasurement.
Unit
Unit of measurement in mmol/l (regardless of language setting). Will not be present if it is a
mismeasurement.
Failure
Error identifier (see Error messages section).
DeviceNo
Serial Number of ChloroChek (alphanumeric).
CheckSum
Checksum of previous values in this line (see Checksum section).
Message
Human-readable message in selected language (not present if there is no notification).
Example of a Single Measurement
In the case of a single measurement the entire ticket is published in one piece when the measurement has been
completed; the value BatchID does not apply.
<Ticket class="SAMPLE" serialno="3400140127" versions="CHLOROCHEK;Main:V1.80;COM:V1.6">
<Measurement>
<SampleId>01</sampleId>
<DateTime>2016-12-22T12:51:30</DateTime>
<Value>21</Value>
<Unit>mmol/l</Unit>
<DeviceCode>3400140127</DeviceCode>
<CheckSum>1fe1219fa897b340c91407c191af319d</CheckSum>
</Measurement>
</Ticket>
DIGITAL DATA MANAGEMENT
5.5 XML Format (continued)
SECTION 5
39