Buffer Solution, Gelatin Solution, And Standard Solution - Elitech ChloroChek Chloridometer 3400 User Manual

Table of Contents

Advertisement

SECTION 2
SAFETY AND HANDLING INFORMATION

2.2 Buffer Solution, Gelatin Solution, and Standard Solution

SS-251 100 mmol/L NaCl/H
Composition:
Container:
Nominal Volume:
Storage and Shelf Life:
The 100 mmol/L NaCl/H
O Standard Solution is not dangerous according to EC Directive 67/548/EEC.
2
NOTE:
The 100 mmol/L NaCl/H
O Standard Solution consists of chemicals. All precautionary measures and regulations
2
must be followed (do not swallow, do not taste, always wear gloves, etc.).
The ampules have a blue dot colored breaking ring. The ampule can be opened manually by breaking it off at this
point. Follow all safety precautions for the handling of glass (splintering, breakage, etc.).
Buffer Solution and Gelatin Solution (Working Solution)
The Buffer Solution and Gelatin Solution are hazardous materials according to EC Directive 67/548/EEC.
Information of safety and handling can be found in the attached safety data sheets.
The shelf life of a fresh Working Solution (a vial of Buffer Solution plus 20 drops of Gelatin Solution), is maximum
of 14 days when kept in a sealed opaque vessel and stored at 25-46.4°F, ( 2˚C to 8˚C). Working Solution (which
contains silver chloride) that is to be used for measuring purposes should be used as soon as possible since it can
react photochemically with light (reduction of the precipitated silver chloride into elementary silver). It is
recommended that a new working solution is prepared for each measurement series. The on-board stability of
the working solution is 24 hours at room temperature.
SS-248BS Buffer Solution:
Composition:
Container:
Nominal Volume:
Storage and Shelf Life:
SS-248GS Gelatin Solution:
Composition:
Container:
Nominal Volume:
Storage and Shelf Life:
Read the Safety Data Sheets (SDS) for Buffer Solution and Gelatin Solution.
CAUTION!
Disposal of all waste material should be in accordance with local, state and Federal regulatory requirements.
The reagent set contains acetic acid and nitric acid. If discharged into the plumbing system, rinse with plenty of
water.
10
O Standard Solution
2
100 mmol/L sodium chloride in distilled water.
Clear glass One Point Cut (OPC) ampule.
1 mL.
Up to the expiration date indicated on the packaging when stored in the ampule at 41-
113°F, (5˚ C to 45 ˚C). Do not use after expiration date.
Once ampule has been opened the shelf life is up to ½ hour at 71.6°F, (22˚ C).
Solution of: diluted acetic acid and nitric acid, pH 1.12.
Screw top plastic vials.
10 mL.
Up to the expiration date indicated on the packaging when stored in the ampule at 41-
113°F, (5˚ C to 45 ˚C). Do not use after expiration date.
Gelatin Solution with pH indicator, pH 5-7.
Screw top dropper bottle plastic vials.
30 mL.
Up to the expiration date indicated on the packaging when stored in the ampule at 41-
113°F, (5˚ C to 45 ˚C). Do not use after expiration date.