NOTE:
If conditioning is unsuccessful or the software 'times out'. The Measurement Electrode, silver anode, and silver
cathode must be cleaned and a new Working Solution prepared. The Measurement Electrode must be replaced if
conditioning continues to fail after multiple cleaning attempts.
Equipment Needed: Deionized water, catch beaker, silver cleaning cloth and microfiber cloth (silver cleaning cloth
and microfiber cloth is provided as RP-487).
1
Place the small pink or green microfiber cloth and the large blue or orange silver polishing cloth near
each other on a lab bench or other hard surface.
2
Remove the electrodes from the ChloroChek.
3
Observe the angle of the measurement electrode tip demonstrated below. When the correct angle has
been achieved, aggressively pull the tips of the Measurement Electrode across the microfiber cloth with
heavy downward pressure against the cloth. Repeat on all sides of the electrode. If done correctly, dark
marks should appear on the cloth.
4
Electrodes may bend; straighten to be parallel if needed.
MAINTENANCE AND TROUBLESHOOTING

4.2 Cleaning the Electrodes