Error Messages And Troubleshooting - Elitech ChloroChek Chloridometer 3400 User Manual

Table of Contents

Advertisement

SECTION 4
MAINTENANCE AND TROUBLESHOOTING

4.1 Error Messages and Troubleshooting

The following is an overview of the individual component groups, identifying the function of each component, its
potential malfunctions, the effects of the malfunctions on the measurement system, and the possible causes of
the malfunction and the procedure for correcting each malfunction. Conditioning cannot help minimize damage
in this case. It is thus of no consequence whether these malfunctions occur during conditioning or during a
sample measurement. Some errors can be remedied directly by the user or an in-house medical equipment
technician; other errors require return of the unit to the manufacturer.
Component
Malfunction
Groups function
Titration
- Fluctuating
electrode Anode
measurement values
or Cathode
during measurement
of the standard
solution or controls.
Measurement
- Conditioning not
electrode
successful.
Working
- Conditioning not
Solution
successful.
Pipette
- Fluctuating
measurement values
during measurement
of the standard
solution or controls.
26
Effect
Possible Cause
- No plausible
- Contaminated
measurement
electrodes.
values.
- Mechanically
defective
electrodes.
- Unable to begin
- Contaminated
the
electrode.
measurement
- Mechanically
series.
defective
electrode.
- Unable to begin
- Incorrect or
the
used solution.
measurement
series.
- No plausible
- Pipette
measurement
defective or
values.
incorrect use of
the pipette.
Solution
- Clean electrodes with
silver cleaning cloth and
rinse with distilled water.
- Replace electrodes and
working solution.
- Clean the measurement
electrode with the silver
cleaning cloth and then
rinse with distilled water.
- Replace the measurement
electrode.
- Use new working solution.
- Inspect the pipette and/or
assess pipetting technique
for errors.
- A sequence of 10 µL water
samples pipetted into a
container on a precision
scale can help assess
pipetting accuracy and
consistency.

Hide quick links:

Advertisement

Table of Contents
loading