SECTION 4
MAINTENANCE AND TROUBLESHOOTING
4.2 Cleaning the Electrodes (continued)
5
Dampen the large silver polishing cloth with deionized water. Clean all four sides of the electrode on the
dampened cloth.
6
Rinse the electrode with deionized water.
7. Dry with lint free tissue, i.e. Kimwipes.
8. Again, straighten the electrodes if needed.
9. Install the measurement electrode into the ChloroChek.
10. Prepare a fresh working solution.
11. Lower probes into the working solution.
12. Press Measure; then Ready.
13. Add 10uL of 100mmol/L NaCl standard. If 'conditioning successful' appears, continue to check
instrument prior to use as described in Section 3.4.
14. At the end of the session, press Standby, then Exit.
15. Remove the electrodes: measurement electrode, silver anode, and silver cathode.
16. Rinse with deionized water.
17. Dry very well with lint free tissue, i.e. Kimwipes.
18. Store all three electrodes in individual plastic containers to prevent oxidation.
28