11.4. How to update the instrument

®
Select "Program update" from the service menu. The URYXXON
Relax now
expects the upload of a new firmware. Follow the instructions that come with the
update-file to finalize the update process.

11.5. Control strip

The check mode for control strips allows you to check the correct functioning
of the instrument using special color control strips. You can order these control
strips from MACHEREY-NAGEL.
Choose the round button Control strip in the check mode measurement menu
(Display 25). The instrument asks for insertion of the colored control strip (Dis-
play 33). The measurement of the control strip starts automatically as soon as
the strip is placed on the sled, if the autostart is activated. Once the measure-
ment has been completed, the measured values are immediately displayed
and / or printed. The result of the measurement is not stored. The values are so
called remission values for the pre-colored test pads on the control strip. These
values have to be compared to a set of standard values from the control strip
package instructions. For additional information on the control strips or in case
the values do not fit the should values in the control sheet, please consult the
package insert, or contact MACHEREY-NAGEL directly.
ATTENTION: Do not touch the test fields of the control strip with your fingers.