Update & Backup; System Reset; Battery; Service - Macherey-Nagel NANOCOLOR Advance Operating Instruction

Hide thumbs Also See for NANOCOLOR Advance:
Table of Contents

Advertisement

System
13.2 Update & Backup

13.3 System reset

13.4 Battery

13.5 Service

13.6 Calibration

Clicking
opens a window with the options Firmware update,
Method update and Backup.
We recommend to make a data backup before each update.
To save customer-specific special methods, we recommend
exporting them to an external storage medium, e.g. a USB
stick.
Download
the
folder
MACHEREY-NAGEL website. Unzip the folder and follow the in-
structions in the update manual.
A system reset sets the device back to the default setting. All
data, special methods and custom settings will be deleted. We
suggest that you backup all saved data, special methods, and
the user-defined settings beforehand (see 13.2 Update & Backup,
page 33).
After pressing
, a dialogue with a confirmation prompt will
open. Confirming with Yes will reset the instrument to the default
settings.
Clicking the Battery icon opens a window with information about
the internal lithium-ion battery of the device. Here information
about capacity, status and voltage are displayed, among other
information. Have this information available when contacting
MACHEREY-NAGEL Technical Support.
Use of the service menu is protected by a password. This menu
is reserved exclusively for authorized technicians for servicing
purposes.
The spectrophotometer performs a self-test each time it starts to
check the basic calibration. The Zero Calibration as well as the
Turbidity calibration can be performed manually, if required, via
the calibration menu
In each case, the result of the calibration can be saved to an ex-
ternal storage device in the form of a CSV file by clicking the
icon.
Spectrophotometer NANOCOLOR
NOTICE
with
the
update
files
Advance 1.0, 11.2020
®
from
the
33

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents