Display Of Results - Macherey-Nagel URYXXON Relax User Manual

Hide thumbs Also See for URYXXON Relax:
Table of Contents

Advertisement

Manual URYXXON

5.2. Display of results

The sequence number (Seq.No.) as well as the patient identification (ID) will be
displayed with the results.
Positive findings are clearly marked by an asterisk (*) on the printout and on the
display. Additionally, it is possible to enable an acoustic signal on positive fin-
dings. For the evaluation of the Microalbumin test strips, findings are displayed
with an additional statement (Repeat with new sample, normal, abnormal, high
abnormal). Additionally, it is possible to enable an acoustic signal for the findings
"Repeat with new sample", "abnormal" and "high abnormal".
The printout is light-sensitive and may turn yellow when exposed to light during
storage. For archiving purposes the printouts should be kept in a dark place
(patient file) or as a photocopy.
The result displayed may be printed again by pressing
will lead back to the start screen.
80
®
Relax
RESULT
Seq.No: 001
ID: 1234567890
15.02.2019
9:47
*BLD
10 ERY/uL
UBG
NORM
BIL
NEG
PRO
NEG
NIT
NEG
KET
NEG
*GLU
50 mg/dL
pH
5.0
SG
1.000
LEU
NEG
Display 4: Result
RESULT
Seq.Nr: 001
ID: 1234567890
15.02.2019
9:47
ALB
10 mg/L
KRE
10 mg/dL
A:K
*
* Repeat with new
sample
Display 5: Result Microalbumin
Manual URYXXON
. The return panel
®
Relax EN, V5.00 / 04.19

Advertisement

Table of Contents
loading

Table of Contents