System Specifications
Location Requirements
• A fume hood within 8' (2.5 m) of unit or adequate exhaust that pulls a minimum of 125 CFM at point of
connection
• Never install a MARS 6 inside a fume hood
• Provides adequate space for sample handling
• Has a dedicated, grounded outlet no more than 8' (2.5 m) from unit
• Is free from vibration of large equipment
• Designated area with a sturdy workbench capable of handling at least 200 lb (91 kg)
• The system weighs 140 lb (63.5 kg) without vessels
• Allows for easy access to the microwave and room for the door to extend fully at 20" (50.8 cm) and has
adequate space for optional computer or printer
• Adequate bench space
• Footprint of unit: 21" (53.3 cm) W x 25" (63.5 cm) D x 25" (63.5 cm) H
• Recommended counter space: 37" (94.0 cm) W x 28" (71.1 cm) D. Shelves or cabinetry above unit is not
recommended. This provides at least 8" (20.3 cm) of open space on each side and 3" (7.6 cm) of open
space in the rear of the instrument for ventilation
Electrical Requirements
This instrument conforms to Globally Harmonized EN61010-1 Standard for Safety Requirements for Electric
Equipment for Measurement, Control and Laboratory Use Part 1: General Requirements (CAN/CSA-C22.2 No.
1010.1-1992).
• Region Specific
• 60 Hz, 208 – 230 V, 15 A @ 230 V
• Instrument is supplied with a NEMA 6 -15 Plug
• Receptacle, non locking
• 50 Hz, 200 – 240 V, 15 A @ 240 V
• Line voltage must not vary more than 10% of its specified level
• Dedicated, grounded outlet no more than 8' (2.5 m) from unit
• Instrument is supplied with a region-specific power cord or an adequately rated power cord may be used
• Fuse Replacements
• 50/60 Hz, 200 - 240 V: 250 V, 15 A 6.3 mm x 32 mm, fast acting
• Power Output: 1800 watts, 2450 MHz – continuous power available at all power levels to provide more control
for reactions (IEC 705 Method – 1988)
Environmental Requirements
• A fume hood within 8' (2.5 m) of unit or adequate exhaust that pulls a minimum of 125 CFM at point of
connection
• Indoor use only
• Altitude up to 6000' (2000 m)
• A Temperature range of 41 °F (5 °C) to 104 °F (40 °C)
• Vent hose is 8' long with a 3
• Relative humidity range of 10% to 85%, non-condensing
∕
" OD × 2
∕
" ID
1
13
8
16
System Specifications
39