Data Management; View, Print, And Export Data - CEM MARS 6 iWave Manual

Microwave accelerated reaction system
Table of Contents

Advertisement

Data Management

View, Print, and Export Data

View data associated with a specific method.
1. Select the method with the data to view.
2. Select the Information icon.
3. The "Summary" will provide information about the method and the run data will be listed in the order they were
run (newest first).
4. Select the desired run data. Scroll to view the "Run Details" and "Sample Information."
5. Run data can be exported and/or printed.
5.1. To export run data:
• Install a USB in the USB port on the right side of the MARS 6 instrument.
• Select the Export icon to export data.
5.2. If the instrument is equipped with a printer, select the Print icon at top of the screen.
NOTE
To Determine information that will be printed, select "System Settings" > "Settings" > "Run."
Graph - Prints graph
Data - Prints Power and Temperature data at time intervals. Default is "OFF." If this is "On", the printout is
long due to time intervals in seconds.
Report - Activates the selections that are made in the "Data Report" section below, example
"Temperature".