Airyscan Imaging Set Up - Zeiss LSM 800 Standard Operation Protocol

Confocal microscope
Table of Contents

Advertisement

Imaging and Flow Cytometry Core CPOS
J. Airyscan imaging set up
1.
Follow air to oil immersion lens change
procedure in training. Switch to 63x objective.
2.
Select pre-set with prefix '0IFCore_Airyscan...'
or
3.
Click 'Smart setup'
4.
select fluorescent dyes and then click '+' button.
5.
Select 'Airyscan' mode.
6.
click 'OK'
7.
Confirm the following settings in 'Acquisition
Mode' window:
8.
'Frame Size' = 256 x 256 px.
intensity adjustment preventing pre-mature
sample bleaching)
9.
'bit' = Either 8 or 16 but follow carefully in step
[J19] when adjusting intensity parameter.
10. 'Scan Speed' = maximum
11. 'Zoom' > 1.3x
12. Start by only checking √ ONE channel in the
'channels' window.
13. Adjust detection wavelength range to suit your
particular fluorescent dye.
14. Turn on 'Continuous' scan mode
15.
! Important!
Go to 'Airyscan' tab in centre
preview area.
16. Click on 'Display' tab under centre preview area.
17. Turn on 'Auto' checkbox in 'Display' tab and
select 'Min/Max' option for real time intensity
display.
18. Adjust 'Master Gain' between 500V to 850V.
ver. 2.1 (Updated: 11/13/2020)
LSM 800 SOP
(Please refer to steps F26 to F32 for graphical
guidance)
(This is to facilitate
J2
J3
[Refer to Step F9 for graphical guide.]
J15
J17
J-15