ResMed AirFit P10 User Manual page 13

Nasal pillows mask for airmini
Hide thumbs Also See for AirFit P10:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 7
Cleaning the mask pillows, frame and the mask tube
Daily:
1. Clean the mask pillows, frame and mask tube daily. Gently rub in
warm (approximately 86°F or 30°C) water using mild liquid detergent
until visually clean.
2. Rinse well under running water and allow to air dry out of direct
sunlight.
In addition to daily cleaning, each week:
1. Clean the mask frame and the end of the mask tube with mild liquid
detergent using a soft bristle brush. Pay particular attention to the
crevices and cavities.
2. Rinse well under running water and allow to air dry out of direct
sunlight.
Cleaning the headgear
Weekly:
1. Soak the headgear in warm water with a mild liquid detergent.
2. Hand wash the headgear with a soft bristle brush.
3. Thoroughly rinse the headgear under running water.
4. Leave the headgear to air dry out of direct sunlight.
Reprocessing the mask between patients
The AirMini tube, vent module and HumidX/HumidX Plus must be
discarded and replaced with new ones as they cannot be disinfected and
should not be used between other patients.
Reprocess this mask when using between patients. Reprocessing
instructions are available on ResMed.com/downloads/masks.
CAUTION
Visual criteria for product inspection: If any visible deterioration of a
system component is apparent (cracking, discoloration, tears etc), the
component should be discarded and replaced.
English
7