Connecting the SpO2 Sensor
Connecting the SpO
2
Selecting the sensor
The reusable Durasensor DS-100A, 57 20 072 is included in
the original delivery package – see Ordering Information,
page 49.
Other optional sensors mentioned in the tables below are
listed in the Ordering Information, page 49.
Only use the SpO
sensors described here.
2
Strictly follow the Instructions for Use of the sensors.
Incorrect positioning or incorrect use can cause tissue
damage.
Select the sensor according to the following criteria:
— Patient weight
— Agitation and movement of the patient
— Possible application position
— Perfusion of the patient
— Duration of use
The following table is intended to help in selecting the
appropriate sensor. All available sensors with their specific
characteristics are listed.
Sensor type
OXISENSOR I-20
Age group
Infants
Weight of patient
1 to 20 kg
Duration of use
Short and long-term
monitoring.
Check application
site at least every 8
hours.
Patient activity
Limited activity
Preferred measuring
Toe
point
*
Sterilization
Sterile pack
*
in unopened, undamaged packaging
16
Sensor
OXISENSOR D-20
Children
10 to 50 kg
Short and long-term
monitoring.
Check application
site at least every 8
hours.
Limited activity
Finger
Sterile pack
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DURASENSOR
OXISENSOR D-25
DS-100A
Adults
Adults
more than 40 kg
more than 30 kg
Short-term
Short and long-term
monitoring.
monitoring.
Change application
Check application
site at least every 4
site at least every 8
hours.
hours.
Inactive patients
Limited activity
only
Finger
Finger
------
Sterile pack