Set-point Value

The set-point value may be adjusted after selecting the set-point memory and pressing SET.
4 50.
If the set-point value does not need to be changed, press and hold EXIT to resume displaying the well
temperature. To change the set-point value, press SET and then press u or d .
60.
When the desired set-point value is reached, press SET to accept the new value and access the
temperature scale units selection. If EXIT is pressed, any changes made to the set-point are ignored.
S
To change temperature units, see

Scan

The scan rate can be set and enabled so that when the set-point is changed the dry-well heats or cools
at a specified rate (degrees per minute) until it reaches the new set-point. With the scan disabled the
dry-well heats or cools at the maximum possible rate.

Scan Control

The scan is controlled with the scan on/off function that appears in the main menu after the
temperature scale units.
Sc
Off
Press u or d to toggle the scan on or off.
On
Press SET to accept the present setting and continue. Press EXIT to cancel.
S
13
Set-point 4 value in °C
New set-point value
Accept new set-point value
Change the Display
Sc flashes for one second and then the current scan setting displays
Scan function off
Scan function on
Accept scan setting
Units.
Dry-well Calibrator
Operation