Connect a 3-wire power
cable to the 4-screw termi-
nal strip; according to the
voltage level as indicated and
pay particular attention to
the correct live, earth and
neutral terminal connections.
For Model E, HI 931500,
HI 931501 and HI 931500
connect the pH or ORP elec-
trode to the BNC marked
"ELECTRODE INPUT".
For Model T, connect the 2
signal wires of the pH or
ORP transmitter to the ter-
minal marked "IN/OUT
TRANSMITTER" paying par-
ticular attention to the indi-
cated polarity.
Recorder Terminals: these
contacts are the output ter-
minals for connection to a
recorder. The output is from
0 to 20 mA / 4 to 20 mA as
indicated and is proportional
to the measured pH or ORP
value.
PT 100 Terminals: these
contacts connect the PT 100
temperature sensor for au-
tomatic temperature com-
pensation of pH measure-
ment. If temperature com-
pensation is not required,
connect a 110 Ohm, 0.25W
resistor across the terminals
(equivalent to a fixed tem-
perature of 25°C/77°F).

INITIAL PREPARATION

35
FUSE
FUSE
ELECTRODE
INPUT
+
mA OUT
-
PT100