Specifications Hi 931501 And Hi 932500 - Hanna Instruments HI 8510 Series Instruction Manual

Dual output ph controller
Hide thumbs Also See for HI 8510 Series:
Table of Contents

Advertisement

SPECIFICATIONS HI 931501
Range
Resolution
Accuracy
Typical EMC
Deviation
Installation Cat.
Input
Calibration
Temperature
Compensation
Readout
Recorder Output
Set Point Relay
Power Supply
Environment
Weight
SPECIFICATIONS HI 932500
Range
Resolution
Accuracy
Typical EMC
Deviation
Installation Cat.
Input
Calibration
Readout
Recorder Output
Set Point Relay
Power Supply
Environment
Weight
HI 931501
0.00 to 14.00 pH
±0.02 pH
10
Offset: ±2 pH ∆O trimmer
Slope: 80 to 110% Slope trimmer
Fixed or automatic with PT100
from 0 to 100°C
4-digit LCD plus graphic symbols
0 to 20 mA or 4 to 20 mA
Two, Isolated, 2 A, max 240 V,
resistive load, 1.000.000 strokes
110/115 V or 220/240 V; 50/60 Hz
-10 to 50°C (14 to 122°F)
1 kg (2.2 lb.)
HI 932500
-1000 to +1000 mV
10
Slope: 90 to 110% Slope trimmer
4-digit LCD plus graphic symbols
0 to 20 mA or 4 to 20 mA
One, Isolated, 2 A, max 240 V,
resistive load, 1.000.000 strokes
110/115 V or 220/240 V; 50/60 Hz
-10 to 50°C (14 to 122°F)
1 kg (2.2 lb.)
34
0.01 pH
±0.1 pH
±0.2 mA
II
12
Ohm
1 mV
±5 mV
±6 mV
±0.2 mA
II
12
Ohm