Calibrate objective
magnification
Delete objectives
EVOS
M5000 Imaging System User Guide
1.
Go to the Settings  Objective Selection and Calibration, click the Action
button
next to the newly installed objective, then select Calibrate to launch
the Objective Calibration tool.
2.
Mount the EVOS
Calibration Slide in the vessel holder, select the objective
you want to calibrate, then click Calibrate Objective.
3.
Follow the onscreen instructions to calibrate the objective magnification for the
selected objective. For more information, see "Calibrate the objectives", page 76.
After removing an objective from the objective turret, unassign the objective from
the corrsponding turret position.