Configure gas
inputs
102
1.
Go to the Settings tab, then click Incubator to display the Incubator panel.
2.
Click Incubator Setup to open the Incubator Setup window.
3.
Select the appropriate options for the Gas Inputs that reflects your set-up for
the EVOS
™
Onstage Incubator.
•
For Port 1, you may select Air or Premix.
If you select Premix, manually enter the percentage of the CO
reflect the specifics of your set-up.
•
Port 2 is reserved for Nitrogen only.
•
Port 3 is reserved for CO
9.
Click Apply after you have configured the gas connections for each port.
10. Turn on the regulators on the gas tanks. The meters on the regulators show
the tank fill on the right and gas flow on the left.
11. Set the flow on the regulators as follows. Do not exceed 50 psi of pressure.
•
Air: 40–50 psi
•
CO
: 40–50 psi
2
•
Nitrogen: 40–50 psi
only.
2
EVOS
M5000 Imaging System User Guide
™
and O
to
2
2