Thermo Scientific invitrogen EVOS M5000 Manual page 33

Imaging system for fluorescence and transmitted light applications
Hide thumbs Also See for invitrogen EVOS M5000:
Table of Contents

Advertisement

EVOS
M5000 Imaging System User Guide
4.
If needed, click Target again to identify further objects
to include in your count (for example, targets that might
appear different).
Note: For best results, follow these guidelines when identifying target objects:
When selecting objects, circle the entire object and include a slight border
around it.
To include objects of lower intensity in your count, select dimmer objects
during identification.
Circle only one object at a time to help define object size for segmentation.
5.
To distinguish the target from background, click
Backround, then click and drag to draw a circle (red) in
a background area.
6.
To count closely grouped cells that are touching or overlapping as distinct
objects, select from the Split Cells options:
None: Touching or overlapping objects are not counted separately.
Shape: Distinct objects are identified and counted based on shape.
Intensity: Distinct objects are identified and counted based on pixel
intensity.