Seametrics INW Manual page 7

Panel meter reading an inw pt12 sensor
Table of Contents

Advertisement

Power Connection
INW does not supply a power connection. It is up to the client and their engineers to
wire the power connection.
Meter manufacturer power specifications:
AC: 85 - 265 VAC, 50/60 Hz, 15 Watts
Recommended fuse: 1 A, 250V, slow-blow
Wire: Copper with 60° C or 60/75° C insulation
Caution!
Bulkhead and Sensor Wiring
Baud Rate
The panel meter is set to communicate at 19,200 baud. The normal baud rate for the
PT12 is also 19,200. The PT12 can communicate at 38,400, 19,200, or 9600 baud. The
panel meter can communicate at either 19,200 or 9600 baud, but not at 38,400. If your
PT12 has been set to a different baud rate, contact INW for information on how to
reset the baud rate.
Seametrics • 253.872.0284
Observe all safety precautions. To ensure safety
and prevent damage to the meter, wiring should be
done by a qualified technician in accordance with all
local, state, and federal guidelines.
Bulkhead Wiring
Sensor Wiring
PANEL METER FOR INW PT12 SENSOR