Calibration; Loading The Calibration Disk; System Z O - HP 85106D Manual

Network analyzer system
Table of Contents

Advertisement

Loading the
Calibration Disk
SYSTEM Z
o

Calibration

Load the Calibration Kit disk as follows:
1. Insert the standards definition disk and press:
a. [
], {
}, {
DISK
LOAD
CAL KIT *1
b. {
}, {
CAL KIT*1
*file 1
{
} after file 1 is loaded.
*File 2
2. Press [
] again and verify whether the proper files are loaded. The
CAL
HP 8510C softkey field should display the calibration type and the cal
kit tape constants revision number. Remove the disk after the file is
loaded.
3. Press {
} twice if you are ready to calibrate. The program sets the
RESUME
HP 8510C up and places it in LOCAL operation so you can calibrate the
system using the front panel keys.
Press [
] {
} {
MORE
SET Z0
CAL
4. Perform the Calibration. You need the Pull 2-port calibration for
S-parameter test sets, as in the HP 85106D System.
5. On the HP 8510 press [
6. Next, a series of softkey selections appear on the display: {
{
}{
TRANSMISSION
ISOLATION
of softkeys will appear. Connect each device as directed; the HP 8510C
underlines each device label when the measurement is complete.
If desired, refer to the step-by-step TRL calibration procedure in Chapter 2,
"Operation" of this manual.
7. After making all calibration measurements, press the appropriate {
softkey when the last measurement is completed.
8. Store the calibration in a Cal Set Register (1 through 8) by pressing the
accompanying softkey. If an asterisk (*) appears alongside one of the cal
set registers, a calibration set is already stored there.
}.
}. If an asterisk (*) appears next to file 2, press
} [
] [
].
1
x1
]. Then select {
CAL
TRL 2-PORT
} and{
}. When you press one, another set
LINE
HP 85106D mm-Wave System 3-7
Performance Verification
Calibration
}.
}
REFLECTION
}
DONE