1
Introduction
Table 1-1
Frequency
Range
Frequency
Band
NOTE
Getting Started
The HP 85106D mm-wave S-parameter network analyzer is a banded
waveguide system. Table 1-1 lists the waveguide bands available from
33 GHz to 110 GHz. You must add two HP 85104A series test set modules
for each waveguide band. For fully error-corrected measurements, it is
necessary to use a calibration kit in the same frequency band as the test set
modules being used. Refer to "Frequency Bands and Accessories Available"
on page 1-4 in this manual for the specific model numbers required for your
system.
HP 85106D mm-Wave System Frequency Bands
45 MHz - 50 GHz
33 GHz -50 GHz
Coaxial
(option001)
Q-band
WR-22
The HP 85106D mm-wave system arrives with one each of the following
installed in a system cabinet:
•
HP 8510C Network Analyzer
•
HP 85105A mm-Wave Controller
•
HP 83621B Synthesized Source (quantity of two)
•
System Cabinet - 1600 mm Rack, all instruments noted above are
mounted in this cabinet.
•
An HP 85106D system configuration disk is included with your system.
This disk allows you to load the mm-wave instrument and hardware
states into your system. Refer to Chapter 2, "Operation" for more
information about loading your system's configuration.
To complete your mm-wave system, two test set modules must be ordered
for the frequency band of interest. Refer to "Frequency Bands and
Accessories Available" on page 1-4 in this manual.
40 GHz -60 GHz
50 GHz -75 GHz
Waveguide
U-band
V-band
WR-19
WR-15
HP 85106D mm-Wave System 1-1
75 GHz -110 GHz
W-band
WR-10