Getting Started; Introduction - HP 85106D Manual

Network analyzer system
Table of Contents

Advertisement

1

Introduction

Table 1-1
Frequency
Range
Frequency
Band
NOTE

Getting Started

The HP 85106D mm-wave S-parameter network analyzer is a banded
waveguide system. Table 1-1 lists the waveguide bands available from
33 GHz to 110 GHz. You must add two HP 85104A series test set modules
for each waveguide band. For fully error-corrected measurements, it is
necessary to use a calibration kit in the same frequency band as the test set
modules being used. Refer to "Frequency Bands and Accessories Available"
on page 1-4 in this manual for the specific model numbers required for your
system.
HP 85106D mm-Wave System Frequency Bands
45 MHz - 50 GHz
33 GHz -50 GHz
Coaxial
(option001)
Q-band
WR-22
The HP 85106D mm-wave system arrives with one each of the following
installed in a system cabinet:
HP 8510C Network Analyzer
HP 85105A mm-Wave Controller
HP 83621B Synthesized Source (quantity of two)
System Cabinet - 1600 mm Rack, all instruments noted above are
mounted in this cabinet.
An HP 85106D system configuration disk is included with your system.
This disk allows you to load the mm-wave instrument and hardware
states into your system. Refer to Chapter 2, "Operation" for more
information about loading your system's configuration.
To complete your mm-wave system, two test set modules must be ordered
for the frequency band of interest. Refer to "Frequency Bands and
Accessories Available" on page 1-4 in this manual.
40 GHz -60 GHz
50 GHz -75 GHz
Waveguide
U-band
V-band
WR-19
WR-15
HP 85106D mm-Wave System 1-1
75 GHz -110 GHz
W-band
WR-10

Advertisement

Table of Contents
loading

Table of Contents