System Disk - HP 85106D Manual

Network analyzer system
Table of Contents

Advertisement

Table 2-1
Frequency
Range
Frequency
Band
Machine
Dump Files
NOTE
Operation
System Set Up
Procedure
The following procedure describes loading the network analyzer memory
from the system disk.
1. Load the desired Machine Dump file from the system. Insert the disk
into the internal drive of the HP 8510C external disk drive. Press the
following keys:
[
]
DISK
{
STORAGE IS INTERNAL
{
}
LOAD
{
}
MORE
{
MACHINE DUMP
Use the RPG knob to select the file for the desired frequency band
(see Table ).
Press: {
} to start the process.
LOAD FILE
Machine Dump Files Stored On The mm-Wave System Disk
45 MHz - 50 GHz
33 GHz -50 GHz
Coaxial
((option001)
Q-band
WR-22
MD_COAX
MD_SYNWR22
If your system contains an extra test set (HP 85106D Option 001), verify that
only the desired test set is active (active light ON). If unselected test sets are
active, deactivate the test set temporarily by pressing [
Then return to the desired address.
2. When the Machine Dump file is loaded the system begins sweeping the
full frequency range of the current band. Press [
{
} label to see whether the calibration kit for the desired band is
CAL 1
loaded.
3. The system is ready for operation. You may change the instrument state
settings for specific applications.
2-6 HP 85106D mm-Wave System
}
}
40 GHz -60 GHz
50 GHz -75 GHz
Waveguide
U-band
V-band
WR-19
WR-15
MD_SYNWR19
MD_SYNWR15
75 GHz -110 GHz
W-band
WR-10
MD_SYNWR10
] {
}.
LOCAL
TEST SET
] and read the