Selecting A Test Set; Test Set If Switching; Rf Switch Driver Commands - HP 85106D Manual

Network analyzer system
Table of Contents

Advertisement

Test Set IF Switching

RF Switch Driver
Commands
HP 85105A mm-Wave Controller
Controlling Multiple Test Sets
Coaxial Switch Positions With Four Test Sets
New ADDRESS of Test Set
Test Set Switched
21
1
20
2
22
3
23
4

Selecting a Test Set

The active test set is selected via the built-in ability of the HP 8510 to
generate an addressed command to the test set. Each time the HP 8510
ADDRESS of TEST SET function is changed (see the HP 8510 [
menu), the HP 8510 switches the previously addressed test set IF to external
and the newly addressed test set IF to internal. The test set front panel active
indicator shows the test set status. When the test set is active the IF signals
from the test set are applied directly to J11 TEST SET INTERCONNECT.
When the test set is inactive the IF signals appearing at J10 are passed
through to J11 and on to the next test set or to the HP 8510.
The address of the test set can be changed manually from the HP 8510 front
panel by selecting the {
ADDRESS OF TEST SET
address by pressing [
], or it can be changed via program control using the
x1
HP 8510 HP-IB ADDRESS; command. Each test set HP-IB address is set
with address switches on the test set rear panel.
A related feature of the HP 8510 is that when the HP 8510 {
} function is changed, a code sequence is automatically issued over the
SET
HP 8510 system bus to the device at the ADDRESS of RF SWITCH. In the
recommended configuration, this device is an HP 11713A attenuator/switch
driver, which in turn controls one or more HP 33311C coaxial switches. As
shown in Figure on page 6-14, these switches are used to select which test
sets receive the RF output from the network analyzer source. The exact
command issued depends upon the new value of the {
function, also shown in Figure on page 6-14.
6-14 HP 85106D mm-Wave System
Coaxial Switch Port Selected
Switch #1
Switch #2
Port 1
Port 2
Port 1
Port 2
Port 2
Port 1
Port 2
Port 2
LOCAL
} function. Then enter the test set
ADDRESS OF TEST
ADDRESS OF TEST SET
]
}