Operation

Entering parameters

4. If available close the rotor with the rotor lid.
5. The display shows:
Entering parameters
Acceleration curve
The Multifuge X1R offers you a total of 9 acceleration and 10 braking curves with which samples
and gradients can be centrifuged.
After the centrifuge is turned on, the last running profile selected is shown.
1. Press the
The display shows the message
2. Select the desired profile by entering a digit between 1-9 on the numeric pad. Curve number 1 is
the slowest and curve number 9 the fastest.
3. Confirm your entry by pressing
Braking curve
1. Press the
The display shows the message
2. Select the desired curve by entering a digit between 1-9 on the numeric pad. Curve number 1 is the
slowest and curve number 10 the fastest.
3. Confirm your entry by pressing
Preselecting Speed / RCF
1. Press the
The display shows the
to toggle between the two modes.
Multifuge X1R
4-4
Be sure to check all seals before starting any aerosol-tight applications.
See the information in the rotor instruction manual.
key once in order to open the acceleration profile selection menu.
kay twice in order to open the braking curve selection menu.
key.
RPM
or the "RCF"-value depending on the display setting. Press the
Set
acceleration.
.
Set
deceleration:
.
key