Page 1
MTD33 2000 COUNTS DIGITAL MULTIMETER OPERATION MANUAL 1. SAFETY INFORMATION SAFETY SYMBOLS Warning! Dangerous Voltage (Risk of electric shock). Caution! Refer to the user’s manual before using this Meter. Double Insulation (Protection Class Ⅱ). Alternating Current (AC). Direct Current (DC).
Do not operate the meter if the body of the meter or the test lead looks broken. Check the main function dial and make sure it is at the correct position before each measurement. When making current measurements ensure that the circuit is not “live” before opening it in order to connect the test leads.
ELECTRICAL SPECIFICATIONS Accuracies are ± ( % of reading + number in last digit ) at 23 ± 5°C ,≤75% RH. 2.2.1 DC Voltage Range Accuracy Resolution 200mV 0.1mV 2000mV ± ( 1.0%+2 ) 10mV 200V 100mV 500V ± ( 1.2%+2 ) Overload protection: 500V DC or AC rms Impedance: 1MΩ...
2.2.4 Resistance Range Accuracy Resolution 200Ω ± ( 1.0%+5 ) 0.1Ω 2000Ω 1Ω 20kΩ ± ( 1.0%+3 ) 10Ω 200kΩ 100Ω 2000kΩ ± (1.5%+3 ) 1kΩ Overload protection: F0.5A/600V fuse 2.2.5 Diode and Audible continuity test Range Description Test condition Display read Forward DC current approximately...
Page 5
Don’t apply more than 500V DC or AC rms to the input, indication is possible at higher voltage but there is danger of damaging the internal circuit. Use extreme caution to avoid contact with high tension circuits when measuring high voltage. AC Voltage Measurement Connect the black test lead to the "COM"...
Page 6
Read the result from the LCD panel. Note: If the resistance value being measured exceeds the maximum value of the range selected, an over-range indication will be displayed (“1”), select a higher range. For resistance of approximately 1 megohm and above, the meter may take a few seconds to stabilize, this is normal for high resistance readings.
Caution: Dispose the used batteries according to the rules, which are defined by each community. Warning: If an explosion or fire hazard could occur through fitting a battery of the wrong type. Fuse replacement This meter is provided with a F0.5A/600V fuse to protect the resistance, Diode, Audible continuity test and the current measuring circuits which measure up to 200mA, with a F10A/600V fuse to protect the 10A range.