Technical Specifications - ResMed Mirage Swift II User Manual

Nasal pillows system
Hide thumbs Also See for Mirage Swift II:
Table of Contents

Advertisement

Available languages

Available languages

Disposal
The Mirage Swift II Nasal Pillows System does not contain any hazardous
substances and may be disposed of with your normal household refuse.

Technical Specifications

Pressure–Flow Curve
60
50
40
30
20
10
0
4
6
8
Mask Pressure (cm H
Note: As a result of manufacturing variations, the vent flow may vary from the
values listed in the table.
Dead Space Information
Dead space is the empty volume of the pillow sleeve and short tube assembly
up to the swivel.
The dead space of the Mirage Swift II for all pillow sleeve sizes is 91 mL.
Standards
CE designation in accordance with EC directive 93/42/EEC, class IIa.
EN ISO 14971, ISO 5356-1, ISO 10993-1 and EN ISO 17510-2.
Therapy Pressure
4 to 20 cm H
O
2
Resistance
Drop in measured pressure between the swivel and pillow sleeve (average for
three pillow sleeve sizes):
at 50 L/min: 0.4 cm H
at 100 L/min: 1.6 cm H
10
12
14
16
18
O)
2
O
2
O
2
Pressure
(cm H
O)
2
4
8
12
16
20
20
7
Flow
(L/min)
20
29
37
43
49