ResMed Mirage SoftGel User Manual

Nasal mask
Hide thumbs Also See for Mirage SoftGel:

Advertisement

Quick Links

Mirage™ SoftGel
nasal mask
User Guide
English • Français • Español • Português

Advertisement

Table of Contents
loading

Summary of Contents for ResMed Mirage SoftGel

  • Page 1 Mirage™ SoftGel nasal mask User Guide English • Français • Español • Português...
  • Page 2 Mirage SoftGel Rx Only Fitting / Mise en place / Colocación / Colocação Removal / Enlèvement du masque / Para quitarse la mascarilla / Remoção...
  • Page 3 Mirage SoftGel...
  • Page 4 Mask components / Composants du masque / Piezas de la mascarilla / Componentes da máscara Item / Description / Description / Part number / Code Article / Descripción / Descrição produit / Número Artículo / de parte / Código do Item produto Mask frame / Entourage rigide /...
  • Page 5 Item / Description / Description / Part number / Code Article / Descripción / Descrição produit / Número Artículo / de parte / Código do Item produto Complete system / Système complet / 61600 (S) 61602 (L) Sistema completo / Sistema completo 61601 (M) 61603 (LW) Frame system / Entourage complet /...
  • Page 6 Disassembly / Démontage / Desmontaje / Desmontagem...
  • Page 7 Reassembly / Remontage / Montaje / Remontagem...
  • Page 8: Using Your Mask

    English Mirage SoftGel nasal mask Thank you for choosing the Mirage SoftGel. The Mirage SoftGel is the first in ResMed’s ConvertAble Series—offering a choice of two high performance cushions (Mirage SoftGel and Mirage Activa LT) on the one durable frame.
  • Page 9 • At a fixed rate of supplemental oxygen flow, the inhaled oxygen concentration varies, depending on the pressure settings, patient breathing pattern, mask, point of application and leak rate. • The technical specifications of the mask are provided for your clinician to check that they are compatible with the CPAP or bilevel device.
  • Page 10: Reprocessing The Mask Between Patients

    Reprocessing the mask between patients This mask should be reprocessed when used between patients. Cleaning, disinfection and sterilization instructions are available from the ResMed website, www.resmed.com/masks/sterilization/americas. If you do not have internet access, please contact your ResMed representative. Troubleshooting Problem/possible cause...
  • Page 11 Problem/possible cause Solution Ports cap or elbow incorrectly inserted. Remove the ports cap or elbow from the mask, then reassemble according to the instructions. Vent is blocked or partially blocked. If the vent requires cleaning, use a soft bristle brush. Mask leaks around the face Mask not positioned correctly.
  • Page 12 Technical specifications Pressure–flow The mask contains passive venting to protect against rebreathing. As curve a result of manufacturing variations, the vent flow rate may vary. Pressure Flow (cm H (L/min) Mask pressure (cm H Dead space Physical dead space is the empty volume of the mask to the end of the swivel.
  • Page 13: Limited Warranty

    90 days or in the case of disposable masks the maximum use period. This warranty is only available to the initial consumer. It is not transferable. If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components.
  • Page 14 ResMed shall not be responsible for any incidental or consequential damages claimed to have resulted from the sale, installation or use of any ResMed product. Some regions or states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation may not apply to you.
  • Page 15: Mirage Activa™ Lt

    English Please read this information in conjunction with your Mirage SoftGel User Guide when changing your Mirage SoftGel cushion to the Mirage Activa LT cushion. Intended use The Mirage Activa LT channels airflow noninvasively to a patient from a positive airway pressure device such as a continuous positive airway pressure (CPAP) or bilevel system.
  • Page 16: Troubleshooting

    M, L, LW: 6.06”/154 mm (H) x 3.74”/95 mm (W) x 4.6”/117 mm (D) Select ‘Activa’ (if available) or ‘Standard’ as the mask option Mask setting when using the Mirage Activa LT with ResMed devices that options have mask setting options.
  • Page 17 Fitting and assembly / Ajustement et montage / Colocación y montaje / Encaixe e montagem...
  • Page 18 031425, FR 031430, JP 1192859, JP 1197930, NZ 411222, US D486227, US D487510, US D493521, US D493522, US D545960, US D567935. Others pending. Mirage is a trademark of ResMed Ltd and is registered in U.S. Patent and Trademark Office © 2009 ResMed Ltd.