Page 2
D557800, US D561893, US D562729, US D562976. Other designs pending. Mirage, Mirage Swift, AutoSet CS 2 and VPAP Adapt SV are trademarks of ResMed Ltd and Mirage, Swift, AutoSet CS and VPAP are registered in U.S. Patent and Trademark Office.
Page 3
Headgear / Hoofdband / Hovedbånd / Hodeutstyr / Pääremmit Reassembling the Mirage Swift II / In elkaar zetten van de Mirage Swift II / Genmontering af Mirage Swift II / Montering av Mirage Swift II igjen / Mirage Swift II -nenäkanyylien kokoaminen Top buckle / Bovengesp / Øverste spænde /...
Page 4
Note: The headgear does not need to be taken apart for cleaning. Headgear reassembly instructions are provided here as additional information. / N.B.: De hoofdband hoeft voor het schoonmaken niet uit elkaar te worden gehaald. Instructies voor het in elkaar zetten van de hoofdband zijn hier uitsluitend gegeven als aanvullende informatie.
2 and VPAP Adapt SV devices. Medical Information Note: The Mirage Swift II does not contain latex, PVC, or DEHP materials. If you have ANY reaction to any part of the system, discontinue use and consult your sleep therapist. Warnings and Cautions •...
Page 6
Discontinue using the Mirage Swift II if you frequently experience nasal irritation, and consult your physician. • The Mirage Swift II must be cleaned and disinfected/sterilized if it is used between patients. See section “Reprocessing the Mirage Swift II between Patients” for details.
Page 7
Note: The short tube assembly and cap can be connected to either side of the pillow sleeve. See “Disassembling for Cleaning” and “Reassembling the Mirage Swift II” for instructions on how to remove and attach the short tube assembly and cap.
Do not pull on the short tube or the flow generator tubing as they may get damaged. • To disassemble the Mirage Swift II for daily cleaning, follow steps 1–3. • To disassemble the Mirage Swift II for weekly cleaning, follow steps 1–5.
Reassembling the Mirage Swift II CAUTION Components of the Mirage Swift II pillow sleeve assembly (pillow sleeve, frame and clip—see section A on the illustration sheet) are not compatible for use with Mirage Swift components. Combining these components from the two systems may result in rebreathing of exhaled air.
(D-4). 5 Insert the cap into the other seal ring until it clicks into place (D-5). 6 The assembled system should look like this (D-6). Make sure the ResMed logo on the pillow sleeve faces out—away from the headgear.
Dead space is the empty volume of the pillow sleeve and short tube assembly up to the swivel. The dead space of the Mirage Swift II for all pillow sleeve sizes is 91 mL. Standards CE designation in accordance with EC directive 93/42/EEC, class IIa.
Ordering Information Part No. Item System 60513 Mirage Swift II Nasal Pillow System - EUR 60514 Mirage Swift II Nasal Pillow System - ANZ 60515 Mirage Swift II Nasal Pillow System - AP...