GE enFlow IV System Manual page 7

Fluid and blood warmer
Hide thumbs Also See for enFlow IV:
Table of Contents

Advertisement

Avoid puncturing the heating surface. If damaged, remove the Warmer from service and replace immediately.
This equipment is not intended for use in an oxygen rich environment (defined as >22% O
No modification of this equipment is allowed.
Do not position the device in a way that makes it difficult to disconnect the device.
Due to highly stable components, microprocessor control, and built-in self-tests, an annual performance check is sufficient.
The steps listed in the enFlow IV fluid/blood warming system operational checklist (Appendix F) should be performed at least once a year, or as required by
your accrediting body.
The enCheck is to be used only with the enFlow Warmer and Controller.
Do not touch the enCheck contact plate surface during or immediately after use since it may be very hot.
The warmer and controller may be a potential biohazard during or after use. Handle and dispose of in accordance with acceptable medical practice and
applicable regulations at the end of their life.
4400-0024-EN Rev A
).