Table 4: Warmer indicator status
Status
Warmer
covers
Ready
Open or closed
Device power up
Closed
In operation
Closed
In operation
Closed
In operation
Closed
In operation
Closed
In operation
Closed
In operation
Closed
Continuous
Closed
operation
Refer to Appendix C for the chart on "Warming system response by temperature. "
Refer to Warnings for additional information.

Cleaning the enFlow IV fluid/blood warming system components

Caution

Do not clean with:
 Ketones (MEK, acetone, etc.)
 Abrasive cleaners
Do not sterilize the Warmer with:
 Steam sterilization (autoclave)
 Dry heat
Do not disinfect or sterilize the Controller.
Do not spray or pour cleaning solutions directly on the Controller.
Do not allow cleaning solutions to accumulate on the Controller.
4400-0024-EN Rev A
Disposable
Power LED
cartridge
indicator
None
Flashing Green
every 3 seconds
In place
Red
½ second
duration
In place
Solid Green
In place
Solid Green
In place
Solid Green
In place
Solid Green
In place
Solid Green
In place
Flashing Red
In place
Solid Green
Temperature
Audible
LED indicator
indicator
Unlit
None
Red
½ second beep
½ second
duration
Flashing Blue
None
Solid Blue
None
Solid or Flashing
None
Blue >30 seconds
Solid Green
None
Solid Yellow
None
N/A-Varies
Continuous
audible bursts
Flashing Red
Continuous
audible bursts
High priority
alarm
- 14 of 40 -
Description
Action required
Warmer unit has power, but
None
is not heating
Successful device power
Observe. If the LED does
up and over temperature
not flash red, replace the
circuit test
Warmer; and contact
Technical Support.
Infusate temperature is
None
< 33 °C
Infusate temperature is
None
≥ 33 °C and < 35 °C
Warmer is unable to
Reduce the flow rate if
heat the infusate within
possible. If there is no
operational range. Infusate
change in operational
temperature is
temperature, consider
< 20 °C and/or the flow
replacing the Warmer and
rate is > 200 mL/min.
contact Technical Support.
Infusate temperature is
None
≥ 35 °C and ≤ 42 °C
Infusate (and/or ambient
Observe. This state whereby
temperature) is > 42 °C
the infusate is > 42 ºC
but less than an Over Temp
should only be entered
condition
periodically during changes
in flow rate or infusate
temperature.
Internal failure in the
Replace the Warmer if this
Warmer
occurs, and contact Technical
Support.
Infusate (and/or ambient
Stop the fluid flow, and slide
temperature) is > 45 °C
the Warmer covers open to
signifying an Over Temp
stop warming. Replace the
condition
Warmer if this occurs, and
contact Technical Support.