Servicing the enFlow IV fluid/blood warming system components

The enFlow IV fluid/blood warming system components have been designed to be durable and long lasting. The system uses current Surface Mount
Technology (SMT) and materials. If a System Fault error occurs, then remove the device from service and refer to the table in Appendix B for proper evaluation
and handling. Do not return the device or place it back into service unless it has been cleaned per the instructions above and evaluated by a trained technician
using the instructions in Appendix E and Appendix F. If the unit stops working properly, contact Customer Service to obtain an Return Goods Authorization
(RGA) number prior to returning the unit to Vital Signs. If damage has occurred to the heating surface, immediately remove it from service.
RGA number
The Technical Support Representative will troubleshoot your product issue with you on the phone. If it is necessary to return a product under warranty, a
replacement unit will be shipped to you within 48 hours. (If the product is no longer under warranty, the Customer Advocacy Representative will discuss
replacement options.) You will be issued a RGA number. You will be instructed to return the product in sufficient packaging to prevent damage in transit,
clearly marking the RGA number on the outside of the box. The return address will be provided to you.
Warmer (Model 100 series)
The Warmer is permanently sealed against fluid ingress and has no user serviceable parts inside.
Controller (Model 121 series)
See "Servicing the enFlow IV fluid/blood warming system components" below for information on user serviceable parts inside the Controller. Check the fuses
located in the outside power entry module if the Controller fails to function. The power cord must be removed to do this.
Replaceable parts for the Controller:
Clock battery (See below for replacement instructions)
Fuse – Bussmann #S500-5-R or equivalent (See below for replacement instructions)
Cover
Screws
Pole clamp screw
Power cord
Warmer mount
The Controller should be subjected to routine safety checks as required by local regulations, (i.e., Earthing Impedance, Leakage Current).

Instructions for replacing the Controller clock battery

Clock Battery Specifications
Cell #
Classification
Chemical System
Designation
Nominal Voltage
Typical Capacity
4400-0024-EN Rev A
CR2032
Lithium Coin
Lithium / Manganese Dioxide (Li/MnO
ANSI / NEDA-5004LC, IEC-CR2032
3.0 Volts
220 mAh
- 16 of 40 -
)