Guidance and Manufacturer's Declaration – Emissions
The enFlow 100 with enFlow 121 is intended for use in the electromagnetic environment specified below. The customer or user of the enFlow 100 with enFlow 121 should
ensure that it is used in such an environment.
Emissions Test
Compliance
RF Emissions
Group 1, Class A
Radiated
CISPR 11
Harmonics
Class A
IEC 61000-3-2
Flicker IEC 61000-3-3
Complies
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The enFlow 100 with enFlow 121 is intended for use in the electromagnetic environment specified below. The customer or user of the enFlow 100 with enFlow 121 should
ensure that it is used in such an environment.
Immunity Test
EN/ IEC 60601
Test Level
ESD
Level 3 (±6kV)
EN/IEC 61000-4-2
Contact
Level 3 (±8kV)
Air
EFT
±2kV Mains
EN/IEC 61000-4-4
±1kV I/Os
Surge
±1kV Differential
EN/IEC 61000-4-5
±2kV Common
Voltage
>95% Dip for
Dips/Dropout
0.5 Cycle
EN/IEC 61000-4-11
60% Dip for
5 Cycles
30% Dip for
25 Cycles
>95% Dip for
5 Seconds
Power Frequency
3 A/m
50/60Hz
Magnetic Field
EN/IEC 61000-4-8
Note: During the 5 second event, the enFlow 100 and 121 power off, but return to normal operation as soon as power is restored.
4400-0024-EN Rev A
Electromagnetic Environment – Guidance
The enFlow 100 with enFlow 121 uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
The enFlow 100 with enFlow 121 is suitable for use in all establishments, other than domestic, and those
directly connected to the public low voltage power supply network that supplies buildings used for domestic
purposes.
Compliance Level
Electromagnetic Environment – Guidance
Level 3 (±6kV)
Floors should be wood, concrete or ceramic tile. If floors are synthetic, the
Contact
r/h should be at least 30%.
Level 3 (±8kV)
Air
±2kV Mains
Mains power quality should be that of a typical commercial or hospital
±1kV I/Os
environment.
±1kV Differential
Mains power quality should be that of a typical commercial or hospital
±2kV Common
environment.
100% Dip for
Mains power quality should be that of a typical commercial or hospital
0.5 Cycle
environment. If the user of the enFlow 100 with enFlow 121 requires
continued operation during power mains interruptions, it is recommended
60% Dip for
that the enFlow 100 with enFlow 121 be powered from an uninterruptible
5 Cycles
power supply or battery.
30% Dip for
25 Cycles
See Note 1
3 A/m
Power frequency magnetic fields should be that of a typical commercial
or hospital environment.
- 23 of 40 -