h. Completely close the covers on the Warmer by pressing down on the cartridge and sliding the covers inward toward each other until they meet. Upon
closing the covers, a short audible beep occurs indicating that the Warmer self-test is being performed and confirms operation of temperature sensors
and alarm indicators. After this process is complete, regulated power is delivered to the Warmer's heating surface, which then begins heating the infusate
through the Disposable Cartridge. Adjust the fluid flow to the desired rate.
i.
Place the IV line in the IV Line Clip in order to prevent it from kinking.
j.
The Warmer is designed to be placed on the bed and/or attached to patient coverings in close proximity to the site of infusion using the cord clip
P/N 980309VS-20. Cushion the patient from the Warmer to aid in the prevention of perioperative peripheral neuropathies or heat-related dermal injury.
k. Do not wrap the Warmer in towels, sheets, blankets or drapes.
l.
Opening the Warmer covers immediately stops the heating but not the flow.
m. To turn off the device, use the switch located at the rear of the Controller.
4400-0024-EN Rev A
IV Line Clip
- 11 of 40 -