5. Insert the enCheck unit into the Warmer. The end of the unit is keyed similar to the Cartridge so it will only fit in the correct orientation. Close the covers.
6. Confirm the enCheck is set to the normal mode.
7. Move the MAINS power switch on the back of the Controller to the ON position. Wait for the thermometer to stabilize, ≈ 30 to 60 sec. assuming all
equipment is close to 20 °C.
8. The temperature on the thermometer should be 40 °C ± 2 °C.
Note
: If the temperature in step 8 above is not 40 °C ± 2 °C, take the Warmer unit out of service. Call Customer Service for an RGA. Please reference the
RGA section in "Servicing the enFlow IV fluid/blood warming system components" for further information regarding returns.