2. Leakage current at the AC power cord
Equipment
Safety analyzer with test lead
enFlow Controller
Purpose
This test is run to check the chassis leakage in microamps. For the purpose of this test, the Controller is the chassis.
Procedure
1. Plug the power cord of the Controller into the safety analyzer.
2. Turn the knob on the safety analyzer to the chassis leakage function.
3. Power the safety analyzer by plugging in and setting the MAINS power to ON.
4. Record the polarity and ground readings for the Controller for both power on and off scenarios for the following configurations:
ENCLOSURE LEAKAGE CURRENT
Normal polarity — normal ground
Reverse polarity — normal ground
Reverse polarity — open ground
Normal polarity — open ground
3. Leakage current of the Warmer to the saline in the IV line
Equipment
enFlow Warmer and Disposable Cartridge
Safety analyzer with ECG leads
Saline
Wire
Extension set with a non-venting cap
4400-0024-EN Rev A
Allowable values of continuous LEAKAGE CURRENTS, in μA.
Normal Condition
Single Fault Condition
Double Fault Condition
Single Fault Condition
Fill with saline
- 29 of 40 -
TYPE BF
100
500
500
500
Wire