ZOLL E SERIES EtCO2 Manual page 12

End tidal carbon dioxide
Table of Contents

Advertisement

E Series - End Tidal Carbon Dioxide (EtCO
How EtCO
is Displayed
2
The E Series unit displays the numeric EtCO
units of mmHg, unless configured for percent or kPa.
Refer to the E Series Configuration Guide (Part No.
9650-1201-01) for instructions on how to configure
alternate units of measure. The unit also displays the
number of breaths per minute, labeled "RR" for
respiration rate. In addition, a capnogram may be
displayed using the Wave 2 softkey.
Displaying the Capnogram Waveform
The E Series unit can display one or two waveforms in
Monitor, Defib, or Manual (AED) mode, as long as the
defibrillator is not charging or ECG analysis is not in
progress. The unit displays only one waveform in Pacer
mode.
Note If you don't see the CO
check the sensor cable connection to the E Series
unit. The CO
box is not displayed if the sensor is
2
not connected to the unit. Once the box is
displayed, after power-on, it remains displayed
even if the sensor is disconnected from the unit.
With EtCO
monitoring, the unit can display a capnogram
2
below the ECG trace for a visual indicator of the
moment-by-moment CO
values. The unit displays the
2
capnogram at half the speed of the ECG display, and
provides 8 seconds of data.
The unit temporarily removes the second waveform from
the display when the user presses the CHARGE,
ANALYZE, or ENERGY SELECT buttons, or the Sync
On/Off softkey. The unit restores the second waveform
to the display 4 seconds after:
a shock is delivered
an analysis is completed, unless the defibrillator is
charging
the last Energy Select button press
Sync mode is turned off
To cycle the display from the capnogram, to the
plethysmograph waveform (if SpO
second waveform displayed, press the Wave 2 softkey.
10
) Option Insert
2
value in
2
display box on the monitor,
2
is installed), to no
2
Use the Zoom softkey from the EtCO
adjust the waveform display size. Numbers shown on the
left side of the capnogram display indicate the scaling.
Physiological Monitoring
The physiological monitoring menu includes the
following softkeys: Param, Wave 2, ID#, Alarms, and 12
Lead.
Param Softkey
When you press the Param softkey, the following
softkeys are displayed: Select, Enter, and Return.
Press the:
Select softkey to scroll among the different available
physiological parameters.
Enter softkey to select the parameter that is
highlighted.
Return softkey to return to the physiological
monitoring menu.
Selecting the EtCO2 parameter causes the following
softkeys to appear: Zero, Disable EtCO2 (or Enable
EtCO2), Zoom, MORE, and Return.
Pressing the:
Disable EtCO2 softkey puts the sensor into the
power saving sleep mode.
Enable EtCO2 takes the sensor out of sleep mode
and turns the heater on for normal operation.
MORE softkey causes the following softkeys to
appear: Average, Comp., RR Filter Disable, (or RR
Filter Enable), MORE, and Return.
Return softkey returns you to the physiological
monitoring menu.
See the next sections for a description of the Zero and
Zoom softkey functions.
Pressing the MORE softkey causes the previous
selections to appear. Pressing the Return softkey returns
the user to the physiological monitoring menu. Refer to
the sections "Average Softkey," "Comp. Softkey," and
"RR Filter Enable/Disable" for descriptions of those
softkey functions.
submenu to
2
9650-1212-01 Rev. E

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents