Turn the rotor clockwise as far as it will go.
6
Finish sliding the cable around the rotor on the back of the autosampler.
Loosen the thumbscrew on the rotor to allow the cable to pass through the
7
clamp, with about 2 mm of cable extending past the clamp.
Figure 8 Position of Cable on Rotor
Verify that there is a small gap between the Z-axis slider and cap at the top of
the Z-drive (approximately 1-3 mm). Adjust the cable a bit if necessary.
8
Figure 9 Z-Drive at Top Position With Gap
Hold the cable flat against the rotor and secure the cable by tightening
thumbscrew on the rotor. The thumbscrew should be as tight as possible using
9
just your fingers.