KROHNE OPTIMASS Series Handbook page 78

Mass flowmeters and 050/051 converters
Table of Contents

Advertisement

Viewing sensor signal conditions – Diagnose Fct. 2.9
Menu 2.9 allows the viewing of eight parameters
Key
This displays the temperature of the measuring tube.
By pressing the → key, the temperature will be indicated.
By pressing the ↵ key, you will be returned to the Function display.
This displays the strain value of the measuring tube stain gauge in ohms (where fitted).
This displays the strain value of the inner cylinder strain gauge in ohms (where fitted).
This displays the current resonant frequency of the sensor. This value is primarily used to
calculate the density of the process fluid.
This value indicates the current to the driver in percent. The heavier the fluid, the higher the
number. Air entrainment also shows up as a higher number.
These display the sensor signal level. Should be as per Menu 5.2.4, and be within 2% of
each other.
This displays the number of serial communication errors between the Front End (sensor
mounted) and the converter electronics since power up. Normally displays 0
Viewing hardware and software versions fitted Fct. 2.10
Menu 2.10 allows the viewing of the hardware and software versions installed in the mass flowmeter.
Key
This displays the software version in the converter MFC 050/051
This displays the hardware version of the converter MFC 050/051
This displays the software version installed in the front end electronics mounted on the
sensor.
Display
Line 1
Fct. 2.9
Fct. 2.9.1
Fct. 2.9.2
Fct. 2.9.3
Fct. 2.9.4
Fct. 2.9.5
Fct. 2.9.6
Fct. 2.9.7
Fct. 2.9.8
Display
Line 1
Fct. 2.10
Fct. 2.10.1
Fct. 2.10.2
Fct. 2.10.3
Installation and Operating Instructions OPTIMASS
Line 2
Diagnose
Strain M. T.
Strain I. C.
Tube Freq.
Drive energy
Sensor A
Sensor B
Comm.Errors
Line 2
Versions
Backend SW
Backend HW
Frontend SW