Fct.
Text
Description and settings
No
2
TEST
Main Menu 2. Test functions
2.1
DISPLAY.
Carry out display test
* SURE (NO). Use the ↑ key to select YES, then press ↵ (Duration of test approx.
2.2
CUR. OUT. 1
Test current output I
* SURE (NO). Use the ↑ key to select YES, then press ↵. Use the ↑ key to select
0 mA
To exit test mode, press the ↵ key at any time.
2.3
CUR. OUT. 2
Test current output 2
see CUR. OUT. 1 above
2.4
CUR. OUT. 3
Test current output 3
see CUR. OUT. 1 Fct. 2.2 above
2.5
PULSE OUT.
Test frequency output
* SURE (NO). Use the ↑ key to select YES ,then press ↵. Use the ↑ key to select desired
pulse width from the list below:
Then press ↵. The system now sends pulses of the required width. To stop the test press
↵ twice.
2.6
FREQ. OUT.
Test frequency output
* SURE (NO). Use the ↑ key to select YES ,then press ↵.
* LEVEL LOW 0 volt DC level will be output from the converter.
* LEVEL HIGH (+ V volts dc)
* 1 Hz
* 10 Hz
2.7
ALARM OUT.
Test alarm output
* SURE (NO). Use the ↑ key to select YES, then press ↵
* LEVEL LOW. 0 Volts is output on the alarm terminal.
* LEVEL HIGH . +24V dc is output on the alarm terminal.
2.8
CONTROL.INP
Test control input
* SURE (NO). Use the ↑ key to select YES, then press ↵.
The actual input level, HI or LO, and the selected functions are displayed see Fct. 3.6.1
End test by pressing the ↵ key.
2.9
DIAGNOSE
Submenu 2.9 Diagnose
2.9.1
TUBE TEMP.
Test temperature
Start with the → key. The temperature in °C or °F is displayed.
Use the ↑ key to display the temperature in °F.
End the test by pressing the ↵ key.
2.9.2
STRAIN M.T.
Test strain measuring tube
Start with the → key. The strain resistance in Ohms is displayed. End the test by pressing
the ↵ key.
2.9.3
STRAIN I.C.
Test strain inner cylinder
Start with the → key. The strain resistance in Ohms is displayed. End the test by pressing
the ↵ key.
2.9.4
TUBE FREQ.
Monitor the primary head frequency
Start test with the → key. End the test with the ↵ key.
2.9.5
DRIVE.ENEGY.
Monitor the primary head's drive level
Start test with the → key. End test with the ↵ key.
2.9.6
SENSOR A
Monitor the amplitudes of sensor A and B
2.9.7
SENSOR B
as percentage of their max. value.
Value should correspond to amplitude setting in Fct. 5.2.4
Start test with the → key. End the test with the ↵ key.
2.9.8
COMM.ERROR
Monitor communication errors
Start test with the → key. Number of communication errors since power on is displayed.
S
End the test with the ↵ key.
2.10
VERSIONS
Submenu 2.10 Versions
BACKEND.SW
Monitor the backend software version
2.10.1
Start test with the → key. End test with the ↵ key.
2.10.2
BACKEND.HW
Monitor the backend hardware version
Start test with the → key. End test with the ↵ key.
2.10.3
FRONTEND.SW
Monitor the frontend software version
Start test with the → key. End test with the ↵ key.
30 sec.). Stop test at any time with the ↵ key.
test signals from the list below.
2 mA
12 mA
0.05 mSec ∗
0.4 mSec
∗ 100.0 mSec
10.0 mSec
Use the ↑ key to select test signals from the list below.
* 100 Hz
* 1000 Hz
Press the ↑ key to switch output to:
To exit test mode, press the ↵ key at any time.
Installation and Operating Instructions OPTIMASS
16 mA
20 mA
22 mA
1.0 mSec
∗ 500.0 mSec
59

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents