700HPPK
Service Information
Safe Operation
To verify safe operation of the Product after maintenance
or repair:
WARNING
To prevent eye damage or personal injury,
use approved eye protection.
1.
Attach a pressure gauge to the hose or the
accessory fitting on the calibration manifold
(700HPPK only). Use any approved (properly-rated)
adapter as required.
2.
Close the upper vent valve (700HPPK) and slowly
pressurize the pump/gauge system to 21 MPa
(3000 psi, 210 bar).
3.
Maintain pressure for 60 seconds, and check for
excessive leaks.
4.
Use the upper vent valve on the calibration manifold
(700HPPK only) to slowly release the pressure.
Pressure can also be vented using the vent valve at
the base of the pump.
If the Product fails this verification, contact Fluke
Calibration for service information. See Contact Fluke
Calibration.
See the 700HPPK Instruction Sheet at
for full specifications and usage instructions.