Fluke -700LTP Instruction Sheet

Low-pressure test pump

Advertisement

Quick Links

Introduction
The Fluke-700LTP Low-Pressure Test Pump is a
portable, dual source of vacuum and pressure. Each
pump incorporates a vacuum and pressure selector
and fine adjustment control. The pump has the
following specifications:
Output pressure: 0 to 45 psi / 0 to 3 bar
Output vacuum: 0 to -13 psi / 0 to -900 mbar
Materials: Bright nickel-plated brass, clear
anodized aluminum
Adjustment: Fine volumetric pressure and
vacuum adjuster
Pressure Relief Valve: Adjustable from 0.7 to 45
psi / 50 mbar to 3 bar
Construction: Nylon, stainless steel, brass, PTFE,
and nitrile
Dimensions: 150 mm x 46 mm diameter
Weight: ~600 grams or 1.3 lbs (pump only)
Box Contents
Fluke-700LTP Low-pressure Test Pump
Two flexible hoses with two 1/4 in BSP female
and two 1/4 NPT female quick-fit connectors
Seal set
Adapter: one 1/4 in. NPT male to 1/8 in. NPT
male.
Instruction Sheet
How to Contact Fluke
To contact Fluke, call one of the following numbers:
USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Visit us on the World Wide Web at:
www.fluke.com
PN 1567810 June 2001, Rev.1, 8/02
Arrow.com.
Downloaded from
© 2001, 2002 Fluke Corporation. All rights reserved. Printed in U.S.A.
Fluke-700LTP
Low-Pressure Test Pump
1
8
A
Release valve. Use to reduce or release pressure in the
system. The rate of pressure reduction depends on the
degree of rotation when opening the valve. Minimal force
is required to seal the system.
B
Fine adjustment control. Turn in or out to increase or
decrease pressure accordingly.
Caution
To avoid damage to the pump, do not wind the fine
adjustment any further when the top of the pump
body is visible.
C
Pressure/vacuum selection. Press the selector as
indicated on the label to engage the desired mode. Make
sure that the release valve
motion) prior to pumping.
D
Quick-fit connectors. Fit the hose
by screwing them into the connectors
knurled nut on the connector fully counterclockwise.
E
Hose.
F
Quick-fit Adapters. 1/4 BSP female & 1/4 NPT female.
Use flat seals with BSP threads. Use PTFE with NPT
threads.
G
Pump Handle. If the pump has not been used for a
while, there may be some resistance on the first stroke.
H
Pressure Relief Valve. The maximum output pressure
can be set by adjustment of the pressure relief value
located inside the main piston. Access to the adjustment
screw is through the hollow-handle retaining screw.
Using a small screwdriver, turn adjustment screw
clockwise to increase pressure, and counterclockwise to
reduce pressure.
®
Instruction Sheet
2
3
4
5
7
4
6
700ltp1f.eps
A
is fully closed (clockwise
E
F
and adapters
D
by turning the
1