Specifications; Standard Specifications - YOKOGAWA SC4AJ Series User Manual

Conductivity sensor
Table of Contents

Advertisement

2.

Specifications

2.1
Standard Specifications
Cable with pin terminals (applicable to FLXA202, FLXA21, FLXA402)
Cable with M3 ring terminals (applicable to FLXA402, SC450G)
Cable with M4 ring terminals (applicable to FLXA202, FLXA21)
Variopin connector (applicable to SA11)
Object of measurement:
Measuring principle:
Cell constant:
Measuring range:
For a cell constant:
For a cell constant:
Temperature Range:
Sterilization for electrode:
Pressure range : For electrode, 0 to 1 MPa
MPa
1.2
1.0
0.8
0.6
0.4
PVDF
0.2
0 10 20 30 40 50 60 70 80
Figure 2.1
-1
0.02 cm
:
-1
0.1 cm
:
For electrode, 0 to 110°C
For holder, see Figure 1
135°C (275°F), within 30 minutes in Steam Sterilization
For holder, see Figure 1
SUS316L
100
T he range of tolerance of holders (option: /PS, /PF, /RS, /RF, /SA1, /SA2, /SB1, /SB2,
/SC1) for temperature and pressure
< 2. Specifications >
Conductivity of solutions
Two-electrode system
-1
-1
0.02 cm
, 0.1 cm
0-0.5 μS/cm to 0-200 μS/cm
0-5 μS/cm to 1 mS/cm
120
140
˚C
F03.ai
IM 12D08F03-02E
10th Edition: Sep. 30, 2019-00
2-1